Limitation of sample volume: How to recover a sample after NTA measurement

Full recovery of samples measured by a nanoparticle tracking analyzer (NTA) often becomes challenging particularly if the samples are intended for further use after measurement.

Figure 1: NTA-cell assembly showing the inlet and outlet ports as well as the flow-through cell channel. The sample is injected into the right inlet, passes the cell channel and leaves the cell assembly on the left outlet.

Abstract

Full recovery of samples measured by a nanoparticle tracking analyzer (NTA) often becomes challenging particularly if the samples are intended for further use after measurement. Furthermore, depending on isolation techniques and purification methods, especially for biological nanoparticles, the yield of samples is often very low in terms of volume and concentration. Here, we describe a convenient way to fully recover samples in order to re-use them entirely for other purposes after they have been measured in the Particle Metrix ZetaView® instrument when equipped with an NTA- cell assembly.

Introduction

The Particle Metrix ZetaView® is widely used in life science research for measuring size, concentration, fluorescence and zeta potential of biological nanoparticles, such as extracellular vesicles, proteins, viruses, prions and other cell-derived nanoparticles (1, 2, 3). Despite well-developed extraction techniques and sophisticated isolation methods, the resulting yield of purified nanoparticles is often very low, particularly if isolated only on the laboratory scale.

For that reason, it is important to sacrifice as small a sample volume as possible for measurements -or even better, to recover the sample once it has been measured.

Using the Particle Metrix ZetaView® in combination with the NTA- cell assembly (not possible with the Z-NTA cell assembly) you can easily recover valuable samples after they have been measured.

Since the NTA- and X-NTA cell assemblies feature a flow-through cell channel system with the inlet and outlet ports located on opposite sides to each other (see Figure 1), you can easily inject your sample on the right side and re-collect it on the left side requiring a minimum of 600µl sample volume.

Figure 1: NTA-cell assembly showing the inlet and outlet ports as well as the flow-through cell channel. The sample is injected into the right inlet, passes the cell channel and leaves the cell assembly on the left outlet.
Figure 1: NTA-cell assembly showing the inlet and outlet ports as well as the flow-through cell channel. The sample is injected into the right inlet, passes the cell channel and leaves the cell assembly on the left outlet.

Methods

Recovering the sample by re-collecting it in a tube

Figure 2: ZetaView set-up with corresponding NTA-cell assembly. The sample is injected via a syringe on the right and collected on the left using an Eppendorf tube.
Figure 2: ZetaView set-up with corresponding NTA-cell assembly. The sample is injected via a syringe on the right and collected on the left using an Eppendorf tube.

After the sample has been measured, it can easily be collected on the left outlet of the cell assembly in an Eppendorf tube or similar vial. Figure 2 shows the complete set-up.

After measurement, the sample is not lost and can be re-collected in an appropriate reaction tube. With this set-up, the sample flow is unidirectional from right to left because the check valves do not allow re-collecting the sample back by drawing it into the syringe.

Recovering the sample by drawing it back into the syringe or re-collecting it in a tube

To have the flexibility to draw the sample back into the syringe after having done the measurement, the check valves (see Figure 1) must be unscrewed in order to allow the sample to go bi-directionally through the flow cell. This set-up is shown in Figure 3.

With the check valves unscrewed from the inlet and outlet ports of the cell assembly, the sample can either be re-collected into a tube at the left side or sucked back into the original syringe at the right side without using the left outlet port. This way of re-collecting the sample is possible if the sample volume that is injected into the instrument does not exceed 1000µl. A larger volume would exceed the capacity of the cell, meaning sample begins to flow out of the left outlet port.

Figure 3: ZetaView set-up as showed in Figure 2 but with the check valves removed. The sample can be injected by a syringe and re-collected in a reaction tube or by drawing it back out of the measurement cell into the syringe after the measurement
Figure 3: ZetaView set-up as showed in Figure 2 but with the check valves removed. The sample can be injected by a syringe and re-collected in a reaction tube or by drawing it back out of the measurement cell into the syringe after the measurement

Conclusion

Re-collecting the entire sample either in a tube or by drawing the sample back into the syringe is of great benefit for researchers struggling with high limited sample volume or very low particle concentrations that needs to be used for further experiments. Therefore, sacrificing as little sample as possible, especially if the particle concentration is so low that the sample needs to be measured undiluted, is a great advantage.  

References

  1. Shu Liu et al. (2017)
    Prions on the run: How extracellular vesicles serve as delivery vehicles for self-templating protein aggregates
    Prion., Mar 4;11(2):98-112. doi: 10.1080/19336896.2017.1306162.
     
  2. Kordelas, L. et al. (2014)
    MSC-derived exosomes: a novel tool to treat therapy-refractory graft-versus-host disease.
    Leukemia, 28, 970- 973 doi:10.1038/leu.2014.41
     
  3. NOURI, M. et al. (2012)
    Effect of Partially Hydrolyzed Kappa-Casein on Physicochemical and Sensory Properties of Heated Milk
    Journal of Dispersion Science and Technology, 33. Jg., Nr. 8, S. 1204-1209.

Get the News

Publications

Technology

High Throughput ZetaView® System vs High Cost of Other Methods

In addition to using Nano Particle Tracking Analysis (NTA) to measurement the size distribution and concentration of EV samples, both Microfluidic Resistive Pulse Sensing (MRPS) and the Single Particle Interference Reflectance Image Sensor (SP-IRIS) methods have been widely used as an alternative means of characterizing EVs. In this note, we relay two specific cases for using the ZetaView® NTA system to achieve relative “high-throughput” analysis of many EV samples, along estimations of throughput for MRPS & SPIRIS methods for an equivalent number of samples; further, we establish realistic estimates for the high cost of ownership for operating MRPS & SP-IRIS systems as a result of the cost of consumables as well as the substantially greater amount of time spent to run the same number of samples.

Read More
Example of an F-NTA measurement of a bacteriophage Phi6 preparation labelled with Sybr™Gold nucleic acid stain (blue curve) and Cellbrite® Fix 640 lipid layer stain (red curve) compared to scatter-based NTA (grey curve). Purity was calculated to be 82% for dsRNA and 85% for lipid layer containing phage particles.
Application

Rapid determination of purity, integrity and titer
of viruses by Fluorescence Nanoparticle Tracking
Analysis (F-NTA) using the bacteriophage Phi6

Determination of the titer of viruses and bacteriophages is an indispensable key technology in virological research and for diagnostic purposes. Depending on the method used, the measurements are either qualitative or quantitative, very time-consuming, and do not provide information about integrity or aggregation behaviour of the virus particles. Nanoparticle Tracking Analysis with the Particle Metrix ZetaView® instrument allows the user to perform a rapid concentration determination of virus particles. Using the bacteriophage Phi6 as an example, we show how titer, purity and integrity of phage particles can be measured quickly and reliably using the fluorescence detection capability of the Particle Metrix ZetaView® instrument.

Read More
Contact Form Sliding Bar
  • Step 1
  • Step 2

What can we do for you?


Your Contact Data

Please let us know how we can reach you. We will get in touch with you shortly.


Search

Cookies

Sie haben die Möglichkeit zu verhindern, dass von Ihnen hier getätigte Aktionen analysiert und verknüpft werden. Dies wird Ihre Privatsphäre schützen, aber wird auch den Besitzer daran hindern, aus Ihren Aktionen zu lernen und die Bedienbarkeit für Sie und andere Benutzer zu verbessern.